Benefits
Uniform illumination with LT(Light Tunnel) Technology
LT delivers accurate and precise results without the need for a reference dye.
* Patient: KR 794703(Korea), US 8139210(U.S.A.), JP 4751821(Japan), CN 1798969(China)
Real 5-Color Multiplexing
Five band-pass filters each for excitation and emission are included for multiplexing, and since no reference dyes are necessary, true five-color multiplexing is possible.
Real-time Analysis
Use the Melting curve analysis module to measure the melting points of amplicons in real-time.
Fast and Accurate Analysis
The high-performance 2D CCD sensor collects information from all 96 wells simultaneously, removing any detection-lag deviations and producing accurate results.
Touch down PCR
Time increments and temperature increment controls allow for touch-down PCR applications.
Use Various Fluorescent Dyes
The light source provides even excitation from 480 to 690 nm, allowing the use of various fluorescent dyes.
Wide Linear Dynamic Range
The dynamic range of detection is wide at over 109.
Intuitive and Convenient Software
A user-centric GUI lets users design and analyze experiments with ease.
Specifications
Physical specifications |
Dimension (mm) |
355 (W) x 540 (H) x 470 (D) |
Weight (kg) |
30 kg |
Sample capacity / size |
96 well plate / 0.2 ml tubes |
Sample volume |
20 ~ 100 ul (50 ul recommended) |
Power consumption |
100 - 240 VAC, 50/60 Hz, 850 Watts |
Operating temperature |
15 ~ 30 °C |
Operating humidity |
20 ~ 80%, no condensation |
Thermo module specifications |
Method Heating / Cooling |
Peltier |
Temperature range |
4.0°C ~ 99.9°C |
Max ramp rate |
2.5°C/sec |
Temperature accuracy |
± 0.3°C |
Temperature uniformity |
± 0.5°C |
Lid temperature |
90 ~ 120°C |
Temperature increment range |
0.1°C ~ 2.0°C |
Time increment range |
1 sec ~ 60 sec |
Computer specifications |
Operating system |
Windows XP & Window7 (32-bit OS only, S/W version 3.54.4 or later) |
Processor speed |
Intel Dual Core E2160 (1.8GHz) or higher |
Memory |
1GB or higher |
Communication port |
USB 2.0 high speed |
Screen resolution |
1280X1024 or higher |
Optical specifications |
Light source |
Short arc lamp (120W) |
Sensor |
16-Bit 2D CCD |
Excitation Filter / Emission Filter |
5 Sets |