ExiPrep™ 16 Plus extracts nucleic acids and purifies protein automatically up to 16 samples. Through optimized experimental protocols, it not only allows to get reproducible results, but also wide compatibility with various types of nucleic acids and protein samples. Furthermore, this instrument has excellent space utilization thanks to its small size.
Simplified usage and reproducible results with built-in nucleic acid extraction and protein purification protocols optimized for various samples and targets
High-purity nucleic acid extraction
Extraction of highly pure nucleic acids by attaching magnetic particles to the bottom of each well equally and rapidly by magnetic blocks attached to the instrument
Increased elution efficiency allowing complete evaporation of alcohols by its heating block
Prevention of contamination
Substantial blocking of cross-contamination, which may occur in other wells by the aerosols generated from the tip, through the sophisticated designs drop-contamination prevention device located at the bottom of the filter tip
Automatic UV sterilization
Sterilization to prevent infection for the users by the samples through the UV lamp installed inside the instrument
Easy-to-use
User-friendly UI capable of checking the experimental progress rate real-time without the need of seperate PC by equipping with a 3.5-inch touch screen on the front of the device
Excellent Precision
Minimized error range, that may occur during the automatic dispensing of the solution, through installment of an adjustable syringe motor up to 0.1 mm capable of stably distributing in a volume of 25 ul ~ 1000 ul for application to various protocols and obtaining of experimental results
Application
Division
Application
University and Business Institute
Nucleic acid extraction and protein purification from various samples
University Hospital Research Institute
Nucleic acid extraction for research and development from blood, tissues and various biological samples
Food companies, restaurants
Nucleic acid extraction required for GMO/food poisoning test in a short time from various foodstuffs
Military, police
Nucleic acid extraction necessary for biologic terrorism causative bacteria and food poisoning causative bacteria
Workflow
Specifications
Dimensions (cm)
32 (W) x 48.7 (H) x 53.5 (D)
Weight (kg)
22
Voltage / Frequency
100 - 240 VAC, 50/60 Hz
Heat block
40 - 90°C
Temperature range
15 - 35°C
Humidity range
20 - 80%, no condensation
Operating system
Standalone
Communications
TCP/IP
User interface
320 x 240 touch screen graphic LCD
UV sterilization
15 minute cycle
* Specifications of this product are subject to change without notice to improve product performance.
Comparisons
ExiPrep™ 16 Plus
ExiProgen™
DNA/RNA Extraction
O
O
Protein Synthesis
X
O
Protein Purification
X
O
TFT touch screen
O
O
Sample
16 samples
16 samples X Number of instrument
UV sterilization
O
O
Heater & magnetic block
O
O
Contamination shield of catridge
O
O
TCP/IP network
X
Numerous instrument can be connected
Cooling system for sample store
X
O
Automatic aliquoting
X
O
Power supply
24 VDC, 2.7A(60W)
24 VDC, 7.5A(180W)
weight
22 kg
27 kg
* Unlike ExiProgen™, ExiPrep™ 16 Plus does not offer storing function.
Experimental Data
1. Genomic DNA from cultured mammalian cell (HeLa)
Sample volume: 1X106 cells
Protocol number: 109
Total prep time: about 1 hr 30 min.
Average yield: 4-8 μg
2. Genomic DNA from bacteria (E.coli)
Sample volume: 1X109 cells
Protocol number: 109
Total prep time: about 1 hr 40 min.
Average yield: 8-12 µg
3. Genomic DNA from tissue (bovine skeletal muscle)
Sample volume: ~ 25 mg
Protocol number: 102
Total prep time: about 1 hr 30 min.
Average yield: Up to 20 µg
4. Genomic DNA from plant tissue (spinach leaf)
Sample volume: ~ 100 mg
Protocol number: 104
Total prep time: about 1 hr
Average yield: Up to 5 µg
5. RNA extraction from cultured mammalian cells (HeLa).