Type |
STX2-Plus |
STX4 |
Liquid Volumes Required for Stable Measurements |
Greater liquid (media) volumes per sample are required to keep the longer sensing regions fully immersed in order to get stable readings with STX2-PLUS.
Example volumes: For Corning 3470, 24-well transwell, minimum 300 µL on the top (apical) and 850 µL on the bottom (basolateral) media volumes are needed for STX2-PLUS. |
Less liquid (media) volumes per sample are required for the STX4 to get stable and accurate readings. Because the sensing regions of the STX4 are shorter and specially coated, the STX4 needs less liquid volumes than the STX2-PLUS. The use of less media with the STX4 saves your resources.
Example volumes: For Corning 3470, 24-well transwell, minimum 150 µl apical media and 500 µl basolateral media volumes are needed for STX4. |
Chloriding (User Maintenance) Required |
Yes, it is important.
When used daily, the electrode needs to be chlorided by immersing the electrode tips in 3-6% sodium hypochlorite for 10 minutes (followed by a rinse with DI water) to ensure proper electrode functionality. |
No, it is not required.
The electrode does not need to be chlorided. (It is also functionally unaffected by chloriding.) A special coating of the electrode’s sensing regions ensures proper functionality. Less maintenance is required by a user, which saves you time. |
User Replaceable Blades (These blades come in direct contact with samples and contain the sensing region.) |
No.
Blades are nondetachable. |
Yes.
You have the option of changing the blades, making the STX4 more cost-effective in the long run. |
Application |
• Ideal for 24-well transwell cell culture plates.
• Used in resistance and millivolts measurements. |
Same as STX2-PLUS |
Direct connectivity with EVOM3 |
Yes |
Yes |
Compatibility with EVOM2, EVOM, Millicell ERS-2, and ERS (using EVM-AC-02-01-01 adapter) |
Yes |
Yes |