Item |
Specifications |
Parallel measurement |
Up to 2 titration stations are connectable. (with optional titration station (and stirrer)) |
Titration methods |
Potentiometric titration |
Measuring range |
pH: 0 to 14,
mV: -2000 to +2000 mV |
Titration mode |
Learn titration, Inflection-point detection, Set-point detection, Set point/Inflection-point detection, Cross-point (F, B, V1, V2) detection, Neutralization values in petroleum products (Oil1, 2)
Other functions: pH-Stat (1, 2), pH measurement, Dispensing, Calculation, pKa measurement |
End-point detection mode |
Auto-detection using a second derivative (with smoothing function), Set-point detection, Cross-point detection |
Number of endpoints detected |
Up to 5 end points with a Learn titration file or linked method files |
Display |
8.4 inch color LCD touch panel |
Displayed content |
Titration conditions, Data per drip, Titration results, Titration curve, Differential curve, List of titration results, Statistics calculation results, Condition settings display, Real-time display of fluid temperature during titration (when thermistor electrode is connected) |
Memory data capacity
(Each titration station) |
Built-in memory: 100 data points
USB flash memory: (≧256MB) Maximum capacity approx. 9,900 data points |
Settings input method |
Key touch input (English guide displayed) |
Sample data |
Sample No.: 2 digits, ID code: 16 alphanumeric characters |
Master file |
Titration execution files: 51 files |
Conditions file |
51 files (method parameters, constant files, control mode files) |
Number of linked condition file |
9 files (up to 5 files in case of titration files linkage) |
Titration control mode file |
Standard files: 51; Neutralization value files in petroleum products: 51;
Manufacturer settings modes (0–11), Blank modes (12–19), User setting mode (20–50) |
Constant file
(settings parameter of concentration calculation constant) |
51 files
K, L (constants), F (factors), M or N (molarity or normality), B (blank mL), S (sample size), T (temperature °C) |
Sample files |
99 files (Sample No., ID code, Master file and Sample size) |
Auto input function |
Function to automatically set titration results to constant parameters (blank, factor, etc.) .
Condition files for automatic input can be designated. |
Calculation function |
Concentration calculation, Statistics calculations, pH automatic compensation |
Branching function |
After measuring the potential prior to titration, branching into maximum 2 condition files |
GLP-compliant function |
- Buret precision check,
- Electrode check,
- Buret use frequency (display/print),
- Electrode maintenance (display/print),
- Reagent level display and alarm
|
Printer |
Built-in thermal printer (Paper roll width: 58 mm) |
External input⁄output |
RS-232C: 1 port (For balance or computer connection),
LAN: 1 port, USB flash memory stick: 1 port |
Power supply |
AC100-240 V±10 % 50/60 Hz 60 VA (MC-3000 only)
A power cable is not included in this product. (*1) |
Dimensions⁄weight |
235(W) × 400(D) × 250(H) mm ⁄ Approx. 6.5 kg (MC-3000 only) |