Two-syringe infusion pump. This feature-laden two-syringe infusion pump combines a broad speed range and holds a wide range of syringe-sizes to meet the requirements of virtually any laboratory application.
Specification :
Accuracy |
±0.35% |
Reproducibility |
±0.05% |
Syringes (Min/Max) |
0.5 µl to 140 ml |
Flow Rate |
Minimum (0.5 µl syringe): 3.06 pl/min Maximum (140 ml syringe, 37.948 mm inner diameter): 215.8 ml/min |
Display |
4.3 WQVGA TFT Color Display with Touchpad |
Non-Volatile Memory |
Stores all Settings |
Connectors |
RS232 - 9Pin D-Sub Connector RS485 - IEEE-1394 6 pos USB - Type B I/O & TTL - 15 pin D-Sub Connector |
Linear Force (Max) |
34 kg (75 lb) @ 100% Force Selection |
Drive Motor |
1.8 degree Stepper Motor |
Motor Drive Control |
Microprocessor with 1/16 microstepping |
Number of Microsteps per one rev.of Lead Screw |
6,400 |
Step Rate - Minimum |
27.5 sec/µstep |
Step Rate - Maximum |
26 usec/µstep |
Pusher Travel Rate - Minimum |
0.36 µm/min |
Pusher Travel Rate - Maximum |
190.80 mm/min |
Power |
100-240 VAC: 50/60 Hz, 50 W, 0.5 A fuse |
Dimensions |
16.51 x 25.4 x 27.94 cm (6.5 x 10 x 11 in) |
Weight |
4.9 kg (10.75 lb) |
Operating Temperature |
4 degree C to 40 degree C (40 degree F to 104 degree F) |
Storage Temperature |
-10 degree C to 70 degree C (14 degree F to 158 degree F) |
Humidity |
20% to 80% RH non condensing |
Mode of Operation |
Continuous |
Classification |
Class 1 |
Pollution Degree |
1 |
Installation Category |
II |
Regulatory Certification |
CE, UL, CSA, CB Scheme, EU RoHS |