Features
1.Glass-made counter window
This will not be worn out within the clean bench work and be enlarged easy to read the volume.
2. Lock mechanism
New material used for 4notch-lock improves its strength, which minimizes the lever shift from the original position.
3. Body
Materials used for pipette body have high tolerance to solvents and drop impact.
4. Plunger
Non corrosive ceramic used in the plunger. (100µL - 10000µL)
5. Seal ring
Seal ring for the Premium is tested dispensing / aspirating 600,000 times. Non-greased plunger makes it easy to maintain. (2µL-1000µL) The test was conducted by Nichiryo.The test result may vary with condition of usage.
6. Nozzle
Nozzle top has highly abrasion tolerant. (2µL-200µL)
7. Hyper Blower system
Hyper blower system improves dispensing efficiency, especially on 2µL and on 10µL dispensing.
Innovation in "Durability" and "Strength"
- Seal ring for the Premium is tested dispensing / aspirating 600,000 times.( 2μL - 200μL )
- Nozzle top is highly abrasion tolerant. ( 2μL- 1000μL )
- Non corrosive ceramic is used for the plunger ( 100μL - 10mL )
Innovation in "Stability" and "Precision"
- Hyper Blower System, longer second-push stroke, improves dispensing efficiency. (2μL,10μL,20μL)
- Easy-Calibration function provides simple lab calibration.
- Hand temperature doesn’t permeate through the body of the pipette which works to avoid affecting volume measurement. ( Patented )
Innovation in "Fatigue-Free" and "Comfort"
- New round shape of the Premium mitigates fatigue and stress from heavy workload.
- Enlarged push button makes it easy to depress.
- White body color mitigates reflection of room light.
Specifications
Cat. No |
Volume Setting |
Volume range(μL) |
Increments(μL) |
Usable tips |
00-NPP-2 |
Digital Setting |
0.1~2 |
0.002 |
BMT2-UT/UTWR, FUT/FUTRB |
00-NPP-10 |
0.5~10 |
0.01 |
BMT2-SSW/SSWR/SSWRB, FSS/FSSRB |
00-NPP-20 |
2~20 |
0.02 |
BMT2-SG/SGR, FSG/FSGRB |
00-NPP-100 |
10~100 |
0.1 |
BMT2-SG/SGR, FSG/FSGRB* |
00-NPP-200 |
20~200 |
0.2 |
BMT2-SG/SGR, FAG/FAGRB |
00-NPP-1000 |
100~1000 |
1 |
BMT2-LG/LGR, LE/LER, FLG/FLGRB FLE/FLERB |
00-NPP-5000 |
1000~5000 |
10 |
BMT2-X/XR |
00-NPP-10000 |
1000~10000 |
10 |
BMT2-Z/ZR |
AC & CV
Cat. No |
Volume(μL) |
Accuracy(%) |
Precision(%) |
00-NPP-2 |
0.2 |
±12.0* |
≤6.0* |
1 |
±5.0 |
≤2.5 |
2 |
±3.0 |
≤1.0 |
00-NPP-10 |
1 |
±4.0 |
≤3.0 |
5 |
±1.0 |
≤1.0 |
10 |
±1.0 |
≤0.5 |
00-NPP-20 |
2 |
±5.0 |
≤3.0 |
10 |
±1.0 |
≤1.0 |
20 |
±1.0 |
≤0.4 |
00-NPP-100 |
10 |
±2.0 |
≤1.0 |
50 |
±1.0 |
≤0.3 |
100 |
±0.8 |
≤0.3 |
Cat. No |
Volume(μL) |
Accuracy(%) |
Precision(%) |
00-NPP-200 |
20 |
±1.0 |
≤0.5 |
100 |
±0.8 |
≤0.3 |
200 |
±0.8 |
≤0.2 |
00-NPP-1000 |
100 |
±1.0 |
≤0.5 |
500 |
±0.8 |
≤0.3 |
1000 |
±0.7 |
≤0.2 |
00-NPP-5000 |
1000 |
±1.0 |
≤0.3 |
2500 |
±0.8 |
≤0.3 |
5000 |
±0.6 |
≤0.2 |
00-NPP-10000 |
1000 |
±2.0 |
≤0.4 |
5000 |
±0.8 |
≤0.3 |
10000 |
±0.4 |
≤0.2 |
- * The AC and CV above are measured with the disposable tips described in the catalogue.
- * The AC and CV at 0.2μL depend much on the sampling skill and the sampling condition where pipettes are used.
Stands
Cat. No |
Storing capacity |
00-MLT-STD2(ABS) |
6pcs |
Cat. No |
Storing capacity |
00-BM-STD(Acrylic resin) |
6pcs |
S-PAC
Cat. No |
Content |
00-SP-AP |
Pipette |
NPP-10(0.5~10μL) |
NPP-100(10~100μL) |
NPP-1000(100~1000μL) |
Tip |
BMT2-SSR(0.5~10μL) |
BMT2-SGR(2~200μL) |
BMT2-LGR(100~1000μL) |
Tube Opener |
NTO-001 (3pcs) |
00-SP-BP |
Pipette |
NPP-20(2~20μL) |
NPP-200(20~200μL) |
NPP-1000(100~1000μL) |
Tip |
BMT2-SGR(2~200μL)×2 |
BMT2-LGR(100~1000μL) |
Tube Opener |
NTO-001 (3pcs) |
00-SP-HL |
Pipette |
NPP-2(0.1~2μL) |
NPP-10(0.5~10μL) |
NPP-100(10~100μL) |
Tip |
BMT2-UTWR(0.1~10μL)×2 |
BMT2-SSR(0.5~10μL) |
BMT2-SGR(2~200μL) |
Tube Opener |
NTO-001 (3pcs) |
Each package contains 3 Nichipet Premiums, 3-4 sample racked tips and 3 tube openers.